Problem with radiation at CP1 after 2010 #88
Comments
BaptisteVandecrux
added a commit
that referenced
this issue
Aug 14, 2022
- Write report to file instead of saving the command prompt output - Swapped ISWR and OSWR at CP1 after 2012 (see #88) - Swapped ISWR and OSWR at LAR1 until 2012 - Implemented SZA and SAA after Ken's code for PROMICE (close #81) - Some more filtering of ISWR and OSWR - Needed to bring tocgen and jaws_tools back to the root folder
Updated after checking for calibration constants #89: blue line is hourly data, orange is daily mean. Albedo is in a more realistic range but still looks suspicious. |
Sign up for free
to join this conversation on GitHub.
Already have an account?
Sign in to comment
Right now I'm inverting the two channes ISWR and OSWR, but the two fluxes have much lower magnitude after 2010.
Maybe wrong calibration coefficients?
The text was updated successfully, but these errors were encountered: