Rigol DS1054Z Waveform Data script not using Channel Offset #14

Open
BillyPocketNC opened this issue Apr 5, 2022 · 0 comments
Open
Assignees

Comments

@BillyPocketNC
Copy link

BillyPocketNC commented Apr 5, 2022

After grabbing some Data from My DS1054, I found that the Waveform Data script is not properly using the channel offset of the oscilloscope
Chanel 1 was moved down 800mA on the screen
image
When I capture the date the offset is -1.2326559
image
Not sure where the actual issue may be.
image

Update

Adjusting the offset to -2.8526559 seems to fix the problem.
The offset seems to be off by 1.62 V