LipSync was working intermittently and would freeze, then stopped working altogether. Remote inspection of the inside of the joystick showed that one of the wires had pulled out of the Dupont connector.

This may have been a result of the wire getting twisted or pulled when adjusting the angle.

Recommend that the Dupont portion of the Stemma QT be removed and the wire soldered directly to the RJ11 Breakout board in the assembly.