Code for main PCB incubator (atMega64)
Code for pcb display Incubator (atMega64)
at tiny1616 code for receiving 2,2kOhm skin sensor data and send them to main microcontroler
Control Mode | Temperature Range |
---|---|
Airway Temperature | 27.0 - 38.5 °C (adjustable) |
Skin Temperature | 34.5 °C - 35 °C (adjustable) |
Humidity | 50 - 90% RH |
Error | Description |
---|---|
Probe Failure | Sensor probe failure (disconnected or cutted) |
Over Temperature | Temperature control increase reach over 37.5 °C |
Temperature Deviation | After short time temperature reach steady state condition and temperature control increase or decrease 1.5 °C from set point, error will triggered |
Power Failure | Input voltage 220VAC/50-60hz not detected |
Fan Failure | fan connector disconnected or cutted |