Expand Up @@ -361,7 +361,9 @@ position of the Bragg axis for the edge energies of 10 metals: Fe, Co,
Ni, Cu, Zn, Pt, Au, Pb, Nb, and Mo.

#. Be sure that all 10 of these elements are actually mounted on the
reference wheel and configured in the ``xafs_ref.contents`` list.
reference wheel and configured in the ``xafs_ref.mapping`` dict.
(They should be. It would be very unusual for any of these foils
to have been removed from the reference wheel.)

#. Run the command

Expand All @@ -380,15 +382,21 @@ Ni, Cu, Zn, Pt, Au, Pb, Nb, and Mo.
the ``calibrate()`` plan.

#. Examine the data in Athena. Make sure E\ :sub:`0` is selected
correctly for all 10 edges. Copy those values into
:file:`edges111.ini` (or :file:`edges3111.ini`). Compute the
angular positions using
correctly for all 10 edges. Copy those values into the first column
of :file:`edges111.ini` (or :file:`edges311.ini`).

.. code-block:: python
.. attention::

dcm.e2a(<energy values>)
It is no longer necessary to compute the angular positions of
the monochromator. Those will be computed from the edge energy
values you edited into the INI file by hand.

and copy those numbers into the :file:`edgeH11.ini` file.
..
Compute the
angular positions using
.. code-block:: python
dcm.e2a(<energy values>)
and copy those numbers into the :file:`edgeH11.ini` file.

#. Run the command

Expand Down