Baxter SIGMA Spectrum Infusion System version 6.05 (model...
Moderate severity
Unreviewed
Published
May 13, 2022
to the GitHub Advisory Database
•
Updated Feb 2, 2023
Description
Published by the National Vulnerability Database
Mar 26, 2019
Published to the GitHub Advisory Database
May 13, 2022
Last updated
Feb 2, 2023
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.